Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131VF7P Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

This recall has been terminated (originally issued January 31, 2019).

Company
Edwards Lifesciences, Llc
Recall Initiated
December 21, 2018
Posted
January 31, 2019
Terminated
November 2, 2023
Recall Number
Z-0764-2019
Quantity
9 units
Firm Location
Irvine, CA
Official Source
View on FDA website ↗

Reason for Recall

These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.

Distribution

Worldwide Distribution

Lot / Code Info

Lot Number: 61176314

Root Cause

Nonconforming Material/Component

Action Taken

An Urgent Product Recall letter was sent via Fed-Ex to affected customers on 12/21/2018. A second communication was sent via Fed-Ex on 01/10/19. The letter identified the affected product, problem and actions to be taken. Customers were instructed to the return any unused units that are currently in their inventory. Once returned, replacement product will be shipped at no charge. Customers were asked to complete the attached acknowledgement form. For questions contact Technical Support at 1-800-822-8837 option 1.