GloCyte RBC Reagent - Product Usage: The GloCyte Automated Cell Counter System is intended for use by trained healthcare professionals in clinical laboratories to provide quantitative determination of fluorescence labeled total nucleated cells and erythrocytes in cerebrospinal fluid collected from adult and pediatric patients. The GloCyte Low and High Level Controls are assayed hematology controls designed to monitor the performance of the GloCyte Automated Cell Counter System. Assayed parameters include total nucleated cells and erythrocytes.

This recall has been terminated (originally issued February 7, 2018).

Company
Advanced Instruments, Llc
Recall Initiated
February 7, 2018
Terminated
September 24, 2019
Recall Number
Z-0640-2019
Firm Location
Norwood, MA
Official Source
View on FDA website ↗

Reason for Recall

RBC Reagent failure. High fluorescent counts during RBC Reagent + Diluent QC check that will result in failed QC check, preventing user from testing patient samples for RBC.

Distribution

US Nationwide distribution in the states of IN, WA, NC, CA, GA, CO.

Lot / Code Info

Part number 111011, Lot # R1513

Root Cause

Device Design

Action Taken

Customers have been sent a recall notification communicating the reason for recall. Once a root cause is identified, recalling firm will notify customers and make available new RBC Reagent lots to allow customers to resume use of reagent in RBC counting.