Eye Cataract Pack, part numbers AMS1359 and AMS1359-1 Product packaged in a convenient manner for use in a general clinical procedure

Company
Windstone Medical Packaging, Inc.
Recall Initiated
March 16, 2017
Posted
April 11, 2017
Terminated
June 6, 2017
Recall Number
Z-1790-2017
Quantity
1,520 kits
Firm Location
Billings, MT

Reason for Recall

AMS custom kits containing Medtronic Covidien Curity Eye Pad are recalled due to the potential for the sterile packaging of the eye pad to be compromised.

Distribution

Nationwide Distribution including CA, CO, FL, GA, IA, KY, MA, MN, MO, ND, NJ, NV, OH, SD, TX, UT, WI, and WY.

Lot / Code Info

Lot numbers:   62980 63106 63387 63825 63826 64186 64490 64584 65937 75122 75329 77010 77089 77480 77664 78265 78312 78499 79042 79211 79944 79945 80709 80994 82316 82537 82538 82931

Root Cause

Nonconforming Material/Component

Action Taken

AMS sent an Urgent Medical Device Recall letter dated March 16, 2017, and provided yellow stickers "Recall Notice" to customers. They are advised to over-label the affected AMS kits at their facilities. Customers were instructed to place sticker in such a way that it would be immediately noticed but should not cover up or conceal any pertinent information such as Pack Number, Pack Name, Lot Number or Expiration Date on a kit. Customers should check inventory and complete the Recall Reply Form. Fax the completed form to Aligned Medical Solutions at 407-804-8460 or email the completed form to vdavis@alignedmedicalsolutions.com. Customers can call Aligned Medical Solutions at 321-527-7714.