11CM ANGLE ATTACHMENT, BLACK MAX

Company
The Anspach Effort, Inc.
Recall Initiated
October 6, 2016
Terminated
July 26, 2019
Recall Number
Z-1972-2017
Quantity
45
Firm Location
Palm Beach Gardens, FL

Reason for Recall

Supplied Directions for Use (DFU) did not contain a recommended service interval.

Distribution

Worldwide Distribution - US Nationwide in the states of HI, DC including PR and the countries of Guam and Japan

Lot / Code Info

All serial numbers shipped to customers; DFU 18-0062; Part No. B-QD11.

Root Cause

Under Investigation by firm

Action Taken

Consignees were sent a recall notice via mail, dated 10/10/2016. In the notice, the firm notifies consignees of the Directions For Use (DFU) correction that will include recommended servicing intervals. All DFUs distributed after 09/28/2016 will have the correct DFU. The recalling firm instructs consignees to review the service intervals provided for product received, forward the notification to necessary parties and/or facilities, and complete the attached Verification Form. Complete Verification Forms are to be sent by fax to 877-546-5069 or by email to Anspach4870@stericycle.com. Consignees are also asked to maintain a copy of the notice for their records. Any questions can be directed to Customer Service at 1-800-327-6887, option 1.