LeMaitre Vascular Reddick Scoop Tip Cholangiogram Catheter (eIFU) REF# e2401-50

Company
Lemaitre Vascular, Inc.
Recall Initiated
February 7, 2017
Terminated
September 20, 2019
Recall Number
Z-1446-2017
Quantity
4673
Firm Location
Burlington, MA

Reason for Recall

The Contains Latex symbol was not printed onto the box label or the sterile (inner) product pouch

Distribution

Worldwide distribution. US Nationwide, AUSTRIA, BELGIUM, FRANCE, GERMANY, ITALY, LUXEMBOURG, Morocco, Saudi Arabia, Singapore, SPAIN, SWITZERLAND, UAE, and UK

Lot / Code Info

Lot Number/Ex p. Date: RST2390 2018-08 RST2393 2018-09 RST2396 2018-10 RST2397 2018-10 RST2398 2018-10 RST2401 2018-12 RST2402 2018-12 RST2418 2019-02 RST2420 2019-03 RST2423 2019-03 RST2428 2019-04 RST2432 2019-04 RST2433 2019-04 RST2437 2019-06 RST2438 2019-06 RST2441 2019-07

Root Cause

Labeling design

Action Taken

LeMaitre Vascular, Inc. issued a Dear Doctor letter, on 2/7/17, to distributors and hospitals via FedEx overnight. The letter contains a form that is requested to be returned to LeMaitre Vascular, Inc. as a record of notification and reconciliation. Any unused product requested to be returned and replaced. Questions concerning this notice contact 781-221-2266 ext. 183.