Hand Foot II, UVB-138. One device per box; device is surrounded by foam inserts inside a corrugated box. For UV phototherapy treatment.

This recall has been terminated (originally issued March 20, 2017).

Company
National Biological Corp
Recall Initiated
January 27, 2017
Posted
March 20, 2017
Terminated
February 8, 2019
Recall Number
Z-1683-2017
Quantity
7 units
Firm Location
Beachwood, OH
Official Source
View on FDA website ↗

Reason for Recall

The wiring may allow the lamps to turn on with the key switch rather than the timer.

Distribution

Worldwide distribution. US states: IA, IL, MI, NY, PA, and WI. Country: Jordan. There are no government accounts for this recall. There are no Canada accounts for this recall. There are no Mexico accounts for this recall.

Lot / Code Info

Serial numbers: 39870, 39873, 39874, 39866, 39867, 39871, 39869

Root Cause

Employee error

Action Taken

Individual patients were contacted via telephone. Replacement devices have been distributed. The distributor was contacted via email. The firm has requested that consignees return the recalled device to their firm to be held under quarantine.