Ascom Telligence Nurse Call System

Company
Ascom (Us) Inc.
Recall Initiated
November 2, 2018
Recall Number
Z-0486-2019
Quantity
57 units
Firm Location
Lakewood Ranch, FL

Reason for Recall

There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.

Distribution

AL, CA, CO, GA, IN, KY, LA, ME, MI, MO, MN, MS, NE, NV, OH, PA, RI, TX UT, VA, WI

Lot / Code Info

Version 5.0 (Build 5.0.4)

Root Cause

Software Manufacturing/Software Deployment

Action Taken

Ascom Partners / Distributors were contacted by e-Mail, USPS Mail and Phone as necessary on 11/02/2018 to determine where the Telligence 5.0 Product is installed.