Sedecal NOVA FA DR System. for diagnostic radiography.

Company
Sedecal S.A. Pol. Industrial Rio De Janiero Calle Pelaya 9-13 Algete Spain
Recall Initiated
July 18, 2018
Recall Number
Z-0488-2019
Quantity
145 units

Reason for Recall

A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient, user, or third parties.

Distribution

US and Canada

Lot / Code Info

Serial Numbers 07360816 through 17291024

Root Cause

Device Design

Action Taken

Initial communication from SEDECAL to all our distributors was performed on 07/18/2018 by email. The letter explained the problem and requested the consignee cease use of the unit if they observe an abnormal displacement in the vertical axis or any suspicious noise during the vertical movement.