RecallDepth

SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009

Company
Baxter Healthcare Corporation
Recall Initiated
September 18, 2018
Terminated
August 19, 2020
Recall Number
Z-0107-2019
Quantity
354 units
Firm Location
Medina, NY

Reason for Recall

If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.

Distribution

NJ

Lot / Code Info

UDI 00085412610900 All Serial Numbers distributed prior to 07/09/2018

Root Cause

Software design

Action Taken

The firm initiated their field correction by email on 09/14/2018. The notice alerted the consignee of the problem and stated that Baxter would be issuing a software update to correct the issue.