HARMONIC ACE Shears + Adaptive Tissue Technology, 5mm Diameter / 36cm Length / Curved, Product Code HAR36, GTIN 10705036014447 Product Usage: The HARMONIC ACE Shears + Adaptive Tissue Technology are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, pediatric, gynecologic, urologic, thoracic, exposure to orthopedic structures (such as spine and joint space) and other open and endoscopic procedures.

Company
Ethicon Endo-Surgery Inc
Recall Initiated
May 3, 2018
Terminated
November 25, 2019
Recall Number
Z-2049-2018
Quantity
146,326
Firm Location
Blue Ash, OH

Reason for Recall

The devices may have been assembled with an internal component that may cause continuous or inadvertent activation of the device.

Distribution

Worldwide distribution - US Nationwide. The devices were distributed to the following foreign countries: Algeria, Angola, Argentina, Armenia, Australia, Austria, Belgium, Bolivia, Bosnia, Brazil, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, French Polynesia, Georgia, Germany, UK, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mayotte, Mexico, Mongolia, Montenegro, Morocco, Netherlands, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Turkey, Ukraine, UAE, Uruguay, Vietnam.

Lot / Code Info

N93923,  N9392K,  N93978,  P9123A,  P9129W,  P91301,  P91394,  P9139N,  P9144R,  P9149J,  P9148K,  P9168K,  P9173R,  P9174K,  P91795,  P91C51,  P91C83,  P91D30,  P91K68,  P91L0H,  P91K69,  P91L0J,  P91L1Y,  P91L6E,  P91L6J,  P9396A,  P9396C,  P9399T,  P93A1L,  P93M0K,  P93M0L,  P93N01,  P93M5Y,  P93N5A,  P93N3X,  P93N3Y,  P93P09,  P93P26,  P93P7H,  P93R10,  P93R4F,  P93R4G,  P93R56,  P93R6V,  P93T0X,  P93T9L,  P93T9M,  P93U17,  P93U91,  P93V0V,  P93V0W,  P93V5T,  P93V02,  P93V03,  P93X4E,  P93X65,  P93X85,  P93X98,  P93X9E,  P93Y47,  P93Y48,  P93Y8X,  P93X2X,  P93Z95,  P93Z4T,  P93Z5X,  P94015,  P9405H,  P9409W,  P9410Z,  P9414T,  P94A5K,  P94A6A,  P94C11,  P94C47,  P94C5R,  P94C8T,  P94D3J,  P94D3K,  P94D5G,  P94D7Z,  P94D9K,  P94E1Z,  P94E20,  P94E3Z,  P94E8W,  P94F1M,  P94F3A,  P94F5T,  P94F6C,  P94F7L,  P94G1G,  P94G1H,  P94G1J,  P94H0J,  P94H0K,  P94H31,  P94H4V,  P94H8J,  P9139V,  P91749,  P9179U,  P91F2X,  P91L1Z,  P93L47,  P93L76,  P93M6T,  P93M6P,  P93N00,  P93N7G,  P93N7H,  P93P2E,  P93P7G,  P93R0Z,  P93R57,  P93R6U,  P93T20,  P93T26,  P93U0P,  P93U5N,  P93U90,  P93V57,  P93W9X,  P93X2G,  P93W9R,  P93X2W,  P94C10,  P94D0M,  P94D0P,  P94J29

Root Cause

Other

Action Taken

Urgent Voluntary Medical Device Recall (Removal) letters were distributed on 5/3/18. The letters included the following instructions for customers: 1. Examine your inventory immediately to determine if you have product lots subject to this recall on hand and quarantine such product(s). 2. Remove the product lots subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product lots subject to this recall have been forwarded to another facility, contact that facility to arrange return. 3. Complete the Business Reply Form (BRF) (Attachment 2) confirming receipt of this notice and fax or email it to Stericycle at 1-866-913-0665 or ethicon5872@stericycle.com within three (3) business days. Please return the BRF even if you do not have the product lots subject to this recall. 4. Keep this notice visibly posted for awareness until all product lots subject to this recall have been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product lots subject to this recall and keep a copy for your records. 5. Customers are required to return all unused HARMONIC ACE+ Shears with Adaptive Tissue Technology product lots subject to this recall that are in their inventory immediately. Only unused HARMONIC ACE+ Shears with Adaptive Tissue Technology product lots subject to this recall returned by August 31, 2018 will be eligible for replacement. Any unused HARMONIC ACE+Shears with Adaptive Tissue Technology product lots subject to this recall returned after August 31, 2018 will not be eligible for replacement. 6. To return unused HARMONIC ACE+ Shears with Adaptive Tissue Technology product lots subject to this recall, photocopy the completed BRF, place it in the box with the subject products, and affix the pre-paid authorized shipping label included with this recall notification letter. Ethicon wi