Trinity Biotech Captia Measles IgM, REF/Product Code 2326060,Test Kit for 96 Determinations, IVD

This recall has been terminated (originally issued September 12, 2016).

Company
Clark Laboratories, Inc. (Dba, Trinity Biotech Usa)
Recall Initiated
July 11, 2016
Posted
September 12, 2016
Terminated
May 9, 2017
Recall Number
Z-2782-2016
Quantity
US: 6 kits, Foreign: 50 kits
Firm Location
Jamestown, NY
Official Source
View on FDA website ↗

Reason for Recall

The product lot failed to meet QC acceptance criteria during stability testing. The Calibrator OD was dropping causing the positive control to be invalid (running high out of range).

Distribution

Distribution to US nationwide including California, and China (Hong Kong).

Lot / Code Info

Lot Number 2326060-043, Expiration Date 31 May 2017

Root Cause

Nonconforming Material/Component

Action Taken

Recall notification letters (dated 7/11/2016) were sent by FedEx on 7/11/2016. Customers were instructed to discontinue use of any remaining inventory (discard on site if any remaining). Customers were also advised that if the test resulted in a valid run, such that all Controls and Calibrator met the Quality Control Criteria as stated in the package insert, the patient results can be reported and a re-evaluation of patient results is not necessary. For questions, please contact Trinity Biotech's Help Desk at 1-800-325-3424 Option #3.