RecallDepth

SURESOUND22US Multipack (US Distribution) containing 2 packs of six devices. SURESOUND22 Multipack (OUS Distribution) containing 2 packs of six devices. For uterine sound

This recall has been terminated (originally issued November 22, 2016).

Company
Hologic, Inc
Recall Initiated
October 28, 2016
Posted
November 22, 2016
Terminated
August 18, 2017
Recall Number
Z-0655-2017
Quantity
600 units (US) ; 24 units (OUS)
Firm Location
Marlborough, MA
Official Source
View on FDA website ↗

Reason for Recall

Breach in sterile barrier compromising product sterility

Distribution

Worldwide Distribution - US (Nationwide) and Internationally to Finland and Austria.

Lot / Code Info

Lot Code US: 16E25RC 16E25RD 16F16R  Lot Code OUS: 16E25RB

Root Cause

Package design/selection

Action Taken

Hologic notified customers in the United States through Certified letter on 10/28/16. The letter contains instructions with the option of returning or destroying the product, provides contact information, and includes a postage paid return card for the consignee to complete and return to Hologic.Questions or require clarification of the response process please contact 1- 800-442-9892 and selection option 6 and then option 2 for surgical product technical support. For questions regarding this recall call 508-263-8920.

More recalls by Hologic, Inc

View all recalls by this company →