RecallDepth

Monaco RTP System; Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Company
Elekta, Inc.
Recall Initiated
November 3, 2016
Terminated
July 16, 2021
Recall Number
Z-0660-2017
Quantity
671
Firm Location
Atlanta, GA

Reason for Recall

Incorrect Dose when using the reset function.

Distribution

AK, AZ, CA, CO, FL, GA, IL, IN, MA, MI, MN, MO, NV, NH, NY, NC, OH, OK, OR, PA, SD, TX, UT, VA, WA, WI, Algeria, Australia, Austria, Brazil, Canada, China, Cuba, Cyprus, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Laos, Malaysia, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Russia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, Venezuela,

Lot / Code Info

Software Builds 5.11 and 5.20

Root Cause

Software design

Action Taken

Important Field Safety Notice 382-01-MON-006 was sent to all affected customers on 11/3/2016. The notice informs users of the specific product and version numbers affected by the issue, and any work around that can be used to avoid the issue. An acknowledgement form was included which is to be returned to Elekta.