COBAS INTEGRA 800 Tina-Quant Hemoglobin A1cDX Gen.2 Hemoglobin A1cDX test Catalog number 04528123160

This recall has been terminated (originally issued November 11, 2016).

Company
Roche Diagnostics Operations, Inc.
Recall Initiated
October 4, 2016
Posted
November 11, 2016
Terminated
June 5, 2017
Recall Number
Z-0396-2017
Quantity
4976
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

.Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results.

Distribution

Nationwide.

Lot / Code Info

Not applicable

Root Cause

Storage

Action Taken

On 10/4/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected Direct Consignees and Distributors via courier. The recall notifications included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Direct Consignee Actions Required " Follow the instructions in the Customer Actions for Prevention/Detection of the Issue section of this Urgent Medical Device Correction (UMDC). " Complete the attached fax form (7184-00-1016) and fax it to 1-844-294-7181. " File this Urgent Medical Device Correction (UMDC) for future reference. Questions Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC. Distributor Actions Required " Ensure proper storage conditions (2-8C) are maintained for the reagents listed in the table above, per their respective method sheet. " If you have not directly shipped any of the products in the table above, you do not need to distribute the enclosed UMDC to your customers. " If you have directly shipped any of the products in the table above, please distribute the enclosed UMDC 16-157 to those customers. " Complete the attached fax form (7185-00-1016) and fax it to 1-844-294-7181. Questions Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this distributor letter or the enclosed UMDC 16-157.

More recalls by Roche Diagnostics Operations, Inc.

View all recalls by this company →