FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Company
Inpeco S.A. Via San Gottardo 10 Lugano Switzerland
Recall Initiated
October 4, 2016
Terminated
May 31, 2019
Recall Number
Z-0587-2017
Quantity
188 units

Reason for Recall

Inpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets.

Distribution

Worldwide Distribution - US including Texas and New York. Foreign Consignees: Germany, Italy, France, Austria and Spain.

Lot / Code Info

Shucko Socket with P/N: 693822160, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 3 meters; Shucko Socket with P/N: 693822170, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 5 meters; Shucko Socket with P/N: 693822190, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 7 meters; Shucko Socket with P/N: 0C00005640, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 2 meters.  The Serial number of the impacted products globally distributed are the following: ACP.009, ACP.024, ACP.039, ACP.054, ACP.069, ACP.084, ACP.100, ACP.115 , ACP.010, ACP.025, ACP.040, ACP.055, ACP.070, ACP.085, ACP.101, ACP.116, ACP.011, ACP.026, ACP.041, ACP.056, ACP.071, ACP.086, ACP.102, ACP.117, ACP.012, ACP.027, ACP.042, ACP.057, ACP.072,ACP.087, ACP.103, ACP.118 , ACP.013, ACP.028, ACP.043, ACP.058, ACP.073, ACP.088, ACP.104, ACP.119, ACP.014, ACP.029, ACP.044, ACP.059, ACP.074, ACP.089, ACP.105, ACP.120, ACP.015, ACP.030, ACP.045, ACP.060, ACP.075, ACP.090, ACP.110, ACP.121, ACP.016, ACP.031, ACP.046, ACP.061, ACP.076, ACP.091, ACP.107, ACP.122 ,ACP.017, ACP.032, ACP.047, ACP.062, ACP.077, ACP.092, ACP.108, ACP.123 ,ACP.018, ACP.033, ACP.048, ACP.063, ACP.078, ACP.093, ACP.109, ACP.124, ACP.019, ACP.034, ACP.049, ACP.064, ACP.079, ACP.094, ACP.110, ACP.125, ACP.020, ACP.035, ACP.050, ACP.065, ACP.080, ACP.096, ACP.111, ACP.126 ,ACP.021, ACP.036, ACP.051, ACP.066, ACP.081, ACP.097, ACP.112, ACP.127 , ACP.022, ACP.037, ACP.052, ACP.067, ACP.082, ACP.098, ACP.113, ACP.128, ACP.023, ACP.038, ACP.053, ACP.068, ACP.083, ACP.099, ACP.114, ACP.129, FLX.103, FLX.0120, FLX.0131, FLX.0137, FLX.0143, FLX.0161, FLX.0166, FLX.0179, FLX.0104, FLX.0124, FLX.0134, FLX.0140, FLX.0152, FLX.0163, FLX.0175, FLX.0108, FLX.0125, FLX.0135, FLX.0141 FLX.0155, FLX.0164, FLX.0176, FLX.0115, FLX.0126, FLX.0136, FLX.0142, FLX.0158, FLX.0165, FLX.0178, AP2.0001, AP2.0004, AP2.0006, AP2.0008, AP2.0010, AP2.0012, AP2.0014, AP2.0003, AP2.0002, AP2.0005, AP2.0007, AP2.0009, AP2.0011, AP2.0013, AP2.0015, FLX.0008, FLX.0133, FLX.0149, FLX.0153, FLX.0159, FLX.0167, FLX.0170, FLX.0174, FLX.0069, FLX.0139, FLX.0150 FLX.0156, FLX.0162, FLX.0168, FLX.0172, FLX.0179, FLX.0132, FLX.0147, FLX.0151, FLX.0157, FLX.0166, FLX.0169, FLX.0173 ,FLX.0121, FLX.0138, FLX.0144, FLX.0145, FLX.0146, FLX.0148, FLX.0176 and FLX.0181.

Root Cause

Employee error

Action Taken

Inpeco sent an Urgent Field Safety Notice dated October 2016 for the Aptio Automation, the Accelerator a3600 and the FlexLab Shucko power cord cables to all affected US and foreign customers. The Urgent Field Safety Notice described the reason for the recalls and actions that should be taken. The firm is requesting all customers to complete and return the Field Correction Effectiveness Check form attached to the Urgent Field Safety Notice letter within 7 days. The Urgent Field Safety Notice letter should be forwarded to affected departments or consignees. To mitigate any possible risks to the laboratory users' health conditions due to the defective power cords Inpeco released a Technical Service Bulletin (TSB) addressed only to Field Service Engineer with the procedure which has to be performed in order to prevent other cases of damaged sockets. The FSE has to check visually the appearance of the Schuko female socket. If it is intact, without any detectable deformation nor melting the FSE has to verify that the terminals are correctly crimped into each single wire. If the female socket presents alterations, the socket has to be replaced with a new one. The affected sockets will be destroyed. Customers with questions were instructed to call (+41) 91 9118 258.