Human Alpha-1-Antitrypsin Kit for use on the SPAPlus analyser, Product #NK034.S. The firm name on the label is The Binding Site Group Ltd., Birmingham, UK. Product is designed for the quantitative in-vitro determination of Alpha-1-antitrypsin in human serum using the SPAPLUS turbidimetric analyser. The measurement of Alpha-1-antitrypsin aids in the diagnosis of several conditions including adult cirrhosis of the liver. In addition, Alpha-1-antitrypsin deficiency has been associated with pulmonary emphysema. This test should be used in conjunction with other laboratory and clinical findings.

This recall has been terminated (originally issued August 23, 2017).

Company
The Binding Site Group, Ltd.
Recall Initiated
August 23, 2017
Terminated
February 21, 2019
Recall Number
Z-0002-2018
Quantity
361 kits
Firm Location
Birmingham United Kingdom
Official Source
View on FDA website ↗

Reason for Recall

The high control for the kit is recovering low and may be out of range for assigned values.

Distribution

Worldwide Distribution - US Distribution to the state of CA., and to the countries of : United Kingdom, Germany, France, Czech Republic, Belgium, Panama, Bulgaria, Iran, Argentina, Turkey Romania, Colombia, and Saudi Arabia.

Lot / Code Info

Lot #400900, Exp. 12/31/2017, UDI #05051700016421

Root Cause

Under Investigation by firm

Action Taken

Binding Site sent an Field Safety Notice dated August 24, 2017, explaining the reason for recall and requesting the direct account and their customer to remove the affected kits from inventory and dispose of them. The direct account and their customers were instructed to circulate the field safety notice to other persons who use the product within their organization. A response form, to be returned via email, was enclosed for completion to indicate the number of kits discarded. For further questions, please call (858) 453-9177.

More recalls by The Binding Site Group, Ltd.