Oncentra External Beam Oncentra Brachy Product Usage: Oncentra is radiation therapy planning software designed to analyze and plan radiation treatment is three dimensions for the purpose of treating patients with cancer.

Company
Elekta, Inc.
Recall Initiated
December 23, 2016
Terminated
July 16, 2021
Recall Number
Z-0987-2017
Quantity
433
Firm Location
Atlanta, GA

Reason for Recall

Cross profile for Varian 60 degree wedge shows "horns."

Distribution

Worldwide Distribution - US Nationwide in the states of AL, CA, DC, NE, NY, ND, OR, PA, RI, SC, TX, and the countries of Angola, Argentina, Armenia, Australia, Austria, Barbados, Belarus, Brazil, Bulgaria, Canada, China Columbia, Curacao, Czech Republic, Denmark, Finland, France, Germany, Greece, Honduras, Hungary, Iceland, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Kenya, Malaysia, Netherlands, Norway, Oman, Peru, Poland, Portugal, Romania, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Yemen, Zambia.

Lot / Code Info

All software builds

Root Cause

Software design

Action Taken

Elekta sent an Urgent Important Safety Notification letter dated December 23, 2016 to affected customers. The notice informs users of the specific product and version numbers affected by the issue and any work arounds that can be used to avoid the issue. The letter included an acknowledgement form to be returned to Elekta.