RecallDepth

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

This recall has been terminated (originally issued February 25, 2016).

Company
Draeger Medical, Inc.
Recall Initiated
February 4, 2016
Posted
February 25, 2016
Terminated
April 1, 2020
Recall Number
Z-0827-2016
Quantity
2553
Firm Location
Telford, PA
Official Source
View on FDA website ↗

Reason for Recall

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

Distribution

US Nationwide Distribution including Puerto Rico.

Lot / Code Info

Catalog #s 8416400 and 8417400

Root Cause

Software design

Action Taken

Draeger Medical, Inc. will send an Urgent Medical Device Recall Letter to consignees in a method that confirms delivery. For consignees that have received new devices with firmware version 1.50 (127 devices) or through device updates (644 devices; reference Recall Z-0436-2016), those devices will be downgraded to firmware version 1.49 and the batteries replaced. Batteries will continue to be replaced for all consignees every six months until a final solution is available.

More recalls by Draeger Medical, Inc.

View all recalls by this company →