Philips X-Ray Systems, Allura Xper with R8.2.16 Product Usage: The Allura Xper FD1O and Allura Xper FD1O/10 is intended for use in cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electrophysiology (EP). The Allura Xper FD2O, Allura Xper FD2O/10 and Allura Xper FD2O/20 is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures.
This recall has been terminated (originally issued October 1, 2015).
- Recall Initiated
- October 1, 2015
- Terminated
- October 8, 2020
- Recall Number
- Z-1066-2016
- Quantity
- 196 devices
- Firm Location
- Andover, MA
- Official Source
- View on FDA website ↗
Reason for Recall
Upon initiating Fluoroscopy the user may encounter a user message Fluoro failed.
Distribution
US Nationwide in the states of : AK, AL, AZ, CA, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, NJ, NY, OH, OR, PA, TN, TX, UT, WI, WV and the country of Canada.
Lot / Code Info
Product Codes: 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058, 722059 systems with release R8.2.16
Root Cause
Software design
Action Taken
Philips sent a Field Safety Notice to all affected customers on October 1, 2015. The notice informs customers of the issue and the actions planned by Philips. Philips Field Service Engineers are to install Software R8.2.16.1 which addresses the issue. For further information or support concerning the issue, consignees should contact their local Philips Representative or call the Technical Support Line at 800-722-9377.