Philips X-Ray Systems, Allura Xper with R8.2.16 Product Usage: The Allura Xper FD1O and Allura Xper FD1O/10 is intended for use in cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electrophysiology (EP). The Allura Xper FD2O, Allura Xper FD2O/10 and Allura Xper FD2O/20 is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures.

This recall has been terminated (originally issued October 1, 2015).

Company
Philips Electronics North America Corporation
Recall Initiated
October 1, 2015
Terminated
October 8, 2020
Recall Number
Z-1066-2016
Quantity
196 devices
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Upon initiating Fluoroscopy the user may encounter a user message Fluoro failed.

Distribution

US Nationwide in the states of : AK, AL, AZ, CA, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, NJ, NY, OH, OR, PA, TN, TX, UT, WI, WV and the country of Canada.

Lot / Code Info

Product Codes: 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058, 722059 systems with release R8.2.16

Root Cause

Software design

Action Taken

Philips sent a Field Safety Notice to all affected customers on October 1, 2015. The notice informs customers of the issue and the actions planned by Philips. Philips Field Service Engineers are to install Software R8.2.16.1 which addresses the issue. For further information or support concerning the issue, consignees should contact their local Philips Representative or call the Technical Support Line at 800-722-9377.