RecallDepth

BD CD64 APC-R700; Analyte Specific Reagent; Model 657701; Lot No. 4344945, 5037611 and 5063945. Hematology - used as analyte specific reagent - contamination leads to possibility of unexpected staining pattern.

Company
Becton, Dickinson And Company, Bd Biosciences
Recall Initiated
October 30, 2015
Terminated
May 27, 2016
Recall Number
Z-0334-2016
Quantity
23 vials
Firm Location
San Jose, CA

Reason for Recall

Three lots of CD64 (MD22) are contaminated with CD4 antibody.

Distribution

Distributed in the states of FL, WA, PA, IL, GA, RI, TX, ND, and IA.

Lot / Code Info

Model 657701; Lot No. 4344945, 5037611, both with expiry date of 31 DEC 2015; and 5063945 with expiry dte of 29 FEB 2016.

Root Cause

Process control

Action Taken

Urgent Product Recall letters were sent out on October 30, 2015 via certified mail to notify customers of the issue.