Etest PIP/TAZO/CON-4 PTC 256 -pack of 30 tests US. Etest is a quantitative technique for determining the antimicrobial susceptibility of Gram-negative and Gram-positive aerobic bacteria.

This recall has been terminated (originally issued December 23, 2015).

Company
Biomerieux 69280 Marcy Letoile France
Recall Initiated
November 19, 2015
Posted
December 23, 2015
Terminated
June 19, 2017
Recall Number
Z-0440-2016
Quantity
1216 units
Official Source
View on FDA website ↗

Reason for Recall

Due to the potential for reporting erroneous results for Piperacillin/Tazobactam; specifically false susceptible results.

Distribution

Worldwide Distribution- US Nationwide including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GS, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV and WY, and the countries of Argentina, Australia, Austria, Brazil, Canada, Chile, Colombia, Czech Republic, Estonia, France, Germany, Hungary, India, Israel, Italy, Japan, Kazakhstan, Kenya, Kingdom of Saudi Arabia, Kuwait, Lebanon, Lithuania, Malaysia, Macedonia, Oman, Netherlands, Pakistan, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russia, Serbia, Singapore, Slavonia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Uruguay, the United Kingdom, Vietnam and Venezuela.

Lot / Code Info

Ref. 412433, Lot numbers: 1004394540, 1003903640, 1003558430, 1003069250, 1002372580 and 1001938220

Root Cause

Process change control

Action Taken

Consignees (Subsidiaries and Distributors) of bioMrieux, impacted by the identified issue, were notified of the FSCA and corrective action (FSCA notification) electronically on November 24th, 2015. Additionally, the bioMrieux United States subsidiary notified US consignees (distributors/end-users) of the FSCA notification via FedEx on December 2nd, 2015.