30inch Dual Port/Single Bladder Disposable Tourniquet Cuff with PLC, Zimmer Surgical, Inc. Tourniquets are intended to be used by qualified medical professionals to exert enough pressure on the arterial blood flow in a limb to produce a bloodless operating field. Single bladder tourniquet cuffs are generally used for operations lasting less than 90 minutes.

This recall has been terminated (originally issued August 6, 2015).

Company
Zimmer Surgical Inc
Recall Initiated
June 24, 2015
Posted
August 6, 2015
Terminated
November 4, 2016
Recall Number
Z-2349-2015
Quantity
330 units
Firm Location
Dover, OH
Official Source
View on FDA website ↗

Reason for Recall

The inside of the folded Instructions For Use (IFU) pamphlet was missing the usage instructions and warnings.

Distribution

Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV & WY and Internationally to Canada.

Lot / Code Info

Item 60707010500, Lot Z000002535 & Item 60707010500, Lot Z000002541.

Root Cause

Labeling False and Misleading

Action Taken

Zimmer sent an Urgent Medical Device Correction letter dated June 30, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the notification, identify the affected product and replace the IFU with the enclosed corrected IFU. Customers were also instructed to complete the attached Response Form and return it via email to CorporateQuality.PostMarket@zimmer.com. Customers with questions were instructed to call 330-364-0989.