EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 C) semi-critical endoscopes.

This recall has been terminated (originally issued June 5, 2015).

Company
Advanced Sterilization Products
Recall Initiated
May 11, 2015
Posted
June 5, 2015
Terminated
July 6, 2016
Recall Number
Z-1715-2015
Quantity
44 units
Firm Location
Irvine, CA
Official Source
View on FDA website ↗

Reason for Recall

Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.

Distribution

Worldwide Distribution-US (nationwide) and the countries of Australia, Singapore, Canada, and Thailand.

Lot / Code Info

5040207003 5040207004 5040308003 5042100008 5042100016 5042100017 5042100018 5042100019 5042100020 5042100021 5042100022 5042100023 5042100024 5042100025 5042100033 5042100034 5042100035 5042100036 5042100037 5042100038 5042100039 5042100040 5042100060 5042100061 5042100063 5042100064 5042100065 5042100066 5042100067 5042110002 5042110003 5042110004 5042110005 5042110006 5042120002 5042120006 5042130009 5042130011 5042140029 5042140030 5042140031 5042140032 5042140038 5042140039

Root Cause

Nonconforming Material/Component

Action Taken

A customer notification letter dated 5/7/15 will be sent to all customers on 5/13/15 to inform them that ASP is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels and subsequently pass the cycle without cancellation if there is a disconnected scope. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the attached Business Reply Card (double-sided) via mail or by fax to Stericycle: 888-879-8144. Customers with questions regarding the business reply card are instructed to contact Stericycle at 888-345-4671 and reference event #7781.