Telluride Percutaneous Rod Inserter Long, Model #7706-1904. The Inserter is a reusable device provided within an aluminum sterilization case along with other orthopedic manual surgical instruments. The case consists of three trays of instruments, and the middle tray contains a quantity of two Inserters. The Inserter is for use as part of a Telluride System pedicle screw procedure.

This recall has been terminated (originally issued February 11, 2015).

Company
Biomet Spine Llc.
Recall Initiated
December 19, 2011
Posted
February 11, 2015
Terminated
March 23, 2015
Recall Number
Z-1089-2015
Quantity
10
Firm Location
Broomfield, CO
Official Source
View on FDA website ↗

Reason for Recall

The Telluride Percutaneous Rod Inserter Long does not contain a feature that locks the Rod in place during use. There is a potential for the Rod to disengage from the Inserter during insertion, and could lead to harm to the patient.

Distribution

Distributed to TX, OH, NE and NY.

Lot / Code Info

LOT Numbers: L561968 and L570545 Product Code: 7703-1600

Root Cause

Device Design

Action Taken

Consignees were informally contacted 12/16/2011 of units in the field requesting retrieval. All units were in Lanx's (Biomet's) possession at the time of the recall initiation on 12/19/2011.