RecallDepth

Baxter Amia Automated PD systems are used in the treatment of adult renal failure patients to perform automated peritoneal dialysis.

This recall has been terminated (originally issued January 7, 2015).

Company
Baxter Healthcare Corp.
Recall Initiated
April 24, 2014
Posted
January 7, 2015
Terminated
July 21, 2017
Recall Number
Z-0932-2015
Quantity
55 units
Firm Location
Deerfield, IL
Official Source
View on FDA website ↗

Reason for Recall

System error 01779 is produced when the battery cannot be charged due to specific voltage differences between the battery and the charger.

Distribution

Nationwide Distribution including GA, MA, NC, NY, PA, and WA.

Lot / Code Info

Product Code: 5C9310

Root Cause

Under Investigation by firm

Action Taken

Baxter sent an IMPORTANT PRODUCT INFORMATION letter dated April 24, 2014 to all affected customers. The peritoneal dialysis provider consignees were instructed to: 1) complete the enclosed customer reply form and return it to Baxter; and, 2) contact Baxter Technical Services at 800-553-6898 should a System Event 01179 occur. Baxter Technical Services is available 24 hours a day, 7 days a week. Clinical questions can be directed to Baxter's Renal Clinical Helpline at 888-736-2543, option 2 (Monday - Friday, 8:00 AM - 5:00 PM CT). General questions regarding the Import Product Information letter can be direct to The Center for One Baxter at 800-422-9837 (Monday - Friday, 8:00 AM - 5:00 PM CT). For questions regarding this recall call 800-422-9837.

More recalls by Baxter Healthcare Corp.

View all recalls by this company →