LANX, REF, PCR8XXX-XXX, LOT, RXONLY. Packaged within aluminum sterilization case. Product Usage: The Durango implant is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2S1.

This recall has been terminated (originally issued December 16, 2014).

Company
Biomet Spine Llc.
Recall Initiated
March 9, 2011
Posted
December 16, 2014
Terminated
January 7, 2015
Recall Number
Z-0220-2015
Quantity
120
Firm Location
Broomfield, CO
Official Source
View on FDA website ↗

Reason for Recall

Biomet Spine, LLC. announces voluntary recall of Durango Anchored ALIF Plate due to the set screw component potentially becoming dislodged from the plate assembly.

Distribution

US Nationwide Distribution in the states of California, Indiana, Louisiana, and Texas,

Lot / Code Info

Lot numbers L532655, L532690, Product code 8503XXXX

Root Cause

Device Design

Action Taken

Biomet notified customers by phone on March 9, 2011 to return the affected product.