LANX, REF, PCR8XXX-XXX, LOT, RXONLY. Packaged within aluminum sterilization case. Product Usage: The Durango implant is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2S1.
This recall has been terminated (originally issued December 16, 2014).
- Company
- Biomet Spine Llc.
- Recall Initiated
- March 9, 2011
- Posted
- December 16, 2014
- Terminated
- January 7, 2015
- Recall Number
- Z-0220-2015
- Quantity
- 120
- Firm Location
- Broomfield, CO
- Official Source
- View on FDA website ↗
Reason for Recall
Biomet Spine, LLC. announces voluntary recall of Durango Anchored ALIF Plate due to the set screw component potentially becoming dislodged from the plate assembly.
Distribution
US Nationwide Distribution in the states of California, Indiana, Louisiana, and Texas,
Lot / Code Info
Lot numbers L532655, L532690, Product code 8503XXXX
Root Cause
Device Design
Action Taken
Biomet notified customers by phone on March 9, 2011 to return the affected product.