Medtronic, Activa PC, Model 37601, Method of Sterilization: Ethylene Oxide, Single Use Only, Rx Only. The Activa PC neurostimulator is a dual-channel device capable of delivering bilateral stimulation. Activa PC contains a non-rechargeable battery and microelectronic circuitry to deliver a controlled electrical pulse to precisely targeted areas of the brain. The device is typically implanted subcutaneously near the clavicle, connected to an extension and leads, which are implanted in the brain.

This recall has been terminated (originally issued August 18, 2014).

Company
Medtronic Neuromodulation
Recall Initiated
July 14, 2014
Posted
August 18, 2014
Terminated
October 6, 2014
Recall Number
Z-2259-2014
Quantity
6
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Medtronic is recalling six Activa PC (model 37601) Implantable Neurostimulators due to the potential for a damaged electrical component during manufacturing.

Distribution

Distributed in the states of: MA, NC, OH, and TX.

Lot / Code Info

Serial numbers: NKM724776H, NKM724782H, NKM724785H,  NKM724790H, NKM724802H, NKM724843H.

Root Cause

Process control

Action Taken

The firm, Medtronic, notified their Consignees on 07/14/2014 via telephone of the recall. Medtronic representative used telephone script to convey the information. The script was directed to Risk Management or Inventory Management. The caller was to inform consignees of the problem and product being recalled. Advised consignees to quarantine the product and provided the Medtronic Device Removal Reply Form to the consignees via e-mail or Fax. The Reply Form included contact information which was to call 1-800-633-8766 in case they needed to contact a Medtronic representative. The completed form is to be faxed back to 1-800-897-3899 or e-mail a PDF to neuro.quality@medtronic.com. If you have any questions, call 763-526-1294.