Coulter DxH Diluent, Catalog No. 628017 Blood cell analysis comprises diluting a whole-blood sample with a solution that functions as a diluent. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types and the measurement of hemoglobin.

This recall has been terminated (originally issued September 16, 2014).

Company
Beckman Coulter Inc.
Recall Initiated
August 25, 2014
Posted
September 16, 2014
Terminated
February 25, 2015
Recall Number
Z-2651-2014
Quantity
46,050 units total (40,078 units in US)
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

Beckman Coulter is recalling Coulter DxH Diluent because it may be associated with elevated platelet (PLT) background counts between 3-7 x 10(3) cells/uL. This will result in failed Daily Checks on the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems.

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to Canada, Chile, Mexico, Panama, and Thailand. .

Lot / Code Info

Lot No. 3510470, 3510480, 3510490, 3510500, 3510510, 3510520, 3510530, 3510540, 3510550, 3510570, 3510580, 3510590, 3510600, 3510610, 3510620, 3510630, 3510690.

Root Cause

Under Investigation by firm

Action Taken

Beckman Coulter sent an Urgent Medical Device Recall letter dated August 25, 2014, to all customers that purchased the Coulter DxH Diluent. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to complete and return the enclosed response form within 10 days. Customers with questions were instructed to contact Beckman Coulter's Customer Support Center at http://www.beckmancoulter.com, or call (800) 526-7694 in the US and Canada.