Microline Surgical ReNew Laproscopic Fenestrated Grasper Forceps Tips (Reusable) Ref 3222

This recall has been terminated (originally issued July 2, 2014).

Company
Microline Surgical, Inc.
Recall Initiated
May 12, 2014
Posted
July 2, 2014
Terminated
August 13, 2015
Recall Number
Z-1987-2014
Quantity
6290 units
Firm Location
Beverly, MA
Official Source
View on FDA website ↗

Reason for Recall

Potential for the Fenestrated Grasper Tip jaw to break when force is applied to the jaw

Distribution

Worldwide Distribution-USA including the states of CA, HI, ID, MA, MI, MS, PA, TN, UT, and TX. and the countries of Belgium, France, Ireland, Germany, Indonesia, Korea, Norway, Saudi Arabia, Switzerland, Turkey, and UK.

Lot / Code Info

Lot Numbers: Lot Number 00115282  Lot Number 00115283  Lot Number 00115284  Lot Number 00115340  Lot Number 00115341  Lot Number 00115501  Lot Number 00115502  Lot Number 00115865  Lot Number 00116277  Lot Number 00116278

Root Cause

Nonconforming Material/Component

Action Taken

Microline Surgical notified Distributors, Hospitals (Customers/End Users) and Suppliers via letter dated May 15, 2014 with labeling, pictures, and Recall Response Forms via UPS Overnight Air, email, and/or fax. Product is requested to be returned and a RMA number will be issued. Contact 978-867-1775 for questions.