Patella 38mm, REF164-0038 Product Usage: UHMWPE component used in a TKA procedure to resurface the patella and serve as a bearing surface for sliding contact with the femoral component

Company
Ortho Development Corporation
Recall Initiated
July 11, 2014
Posted
August 17, 2014
Terminated
October 6, 2014
Recall Number
Z-2254-2014
Quantity
14
Firm Location
Draper, UT

Reason for Recall

Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.

Distribution

Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.

Lot / Code Info

Patella 38mm Part 164-0038, Lot A135844

Root Cause

Packaging process control

Action Taken

Ortho Development notified all affected consignees/distributors of the recall via email on July 11, 2014 which included a list of the affected product (part numbers, lot numbers and quantities) under their control. An RMA number and return shipping label were also included with the email. Emails were followed up with phone calls to verify receipt of the notification.