Fresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/or refurbished Fresenius hemodialysis machines, models 2008K, 2008K2, 2008K@Home and 2008T

This recall has been terminated (originally issued July 11, 2013).

Company
Fresenius Medical Care Holdings, Inc.
Recall Initiated
May 24, 2013
Posted
July 11, 2013
Terminated
October 7, 2019
Recall Number
Z-1693-2013
Quantity
725 units
Firm Location
Waltham, MA
Official Source
View on FDA website ↗

Reason for Recall

Incorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)

Distribution

Nationwide Distribution including the states of AR, CA, FL, GA, KY, LA, MA, MD, NC, NJ, NY, PA, PR, SC, TX and VA.

Lot / Code Info

Part Number: 640919

Root Cause

Nonconforming Material/Component

Action Taken

Fresenius notified customers on May 24, 2013, by telephone and by certified mail with signature confirmation and fax-back form. Customers were instructed to examine their stock immediately to determine whether they have any of the affected parts on hand. If customers have the affected parts, they will be instructed to immediately discontinue use and place all 0-rings in a secure area for return to Fresenius Medical Care North America. Please contact your Fresenius Medical Care Technical Service Team at 1-800-227-2572, (choose option 4, then option 2) for instructions how to return the recalled product.