RecallDepth

FLEXSTENT Femoropopliteal Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT Femoropopliteal SES clinical product is indicated for treatment of atherosclerotic de novo native superficial femoral artery lesions and proximal popliteal lesions.

This recall has been terminated (originally issued June 6, 2013).

Company
Cordis Corporation
Recall Initiated
April 23, 2013
Posted
June 6, 2013
Terminated
June 27, 2014
Recall Number
Z-1510-2013
Quantity
250 units
Firm Location
Miami Lakes, FL
Official Source
View on FDA website ↗

Reason for Recall

Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.

Distribution

Worldwide Distribution - USA (nationwide) and the states of: AZ, CA, CT, FL, GA, IA, IL, LA, MD, MI, MS, NC, NJ, NY, OH, PA, RI, SC, TX, WA, WI, WV, and the countries of: Czech Republic, Italy, Netherlands, and New Zealand

Lot / Code Info

32 Catalog Numbers   FLX-50030-V1, FLX-60030-V1, FLX-70030-V1, FLX-80030-V1, FLX-50040-V1, FLX-60040-V1, FLX-70040-V1, FLX-80040-V1, FLX-50060-V1, FLX-60060-V1, FLX-70060-V1, FLX-80060-V1, FLX-50080-V1, FLX-60080-V1, FLX-70080-V1, FLX-80080-V1, FLX-50100-V1, FLX-60100-V1, FLX-70100-V1, FLX-80100-V1, FLX-50120-V1, FLX-60120-V1, FLX-70120-V1, FLX-80120-V1, FLX-50150-V1, FLX-60150-V1, FLX-70150-V1, FLX-80150-V1, FLX-50200-V1, FLX-60200-V1, FLX-70200-V1, FLX-80200-V1

Root Cause

Package design/selection

Action Taken

Cordis sent an Urgent Medical Device Correction and Potential Removal letter dated April 24, 2013 e-mail to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately identify and set aside all affected products to prevent further use of the affected product, until inspected. Use only product identified with a blue dot next to the lot number on the carton face label, carton end label, and outer pouch label, or product with upgraded sterile barrier system (not yet available, as of the date of this letter), review, complete, sign and return the enclosed Acknowledgement Form and maintain a copy of this notice with the affected product. For questions related to the issue, please contact our FSS/Cordis representative, or please contact Cordis Direct at (800) 781-0282, Monday through Friday from 9:00 AM to 5:00 PM EDT.

More recalls by Cordis Corporation

View all recalls by this company →