DePuy Spine VIPER POLY SCREW 5.5 8 X70MM Tl Product Code: 1867-15-870 Spinal Implant Component.
- Company
- Depuy Spine, Inc.
- Recall Initiated
- November 7, 2012
- Posted
- January 24, 2013
- Terminated
- December 24, 2013
- Recall Number
- Z-0705-2013
- Quantity
- 34 units
- Firm Location
- Raynham, MA
Reason for Recall
Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head
Distribution
Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.
Lot / Code Info
All lots
Root Cause
Component change control
Action Taken
DePuy Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated November 7, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, quarantine the material and return remaining inventory. Customers were advised to complete the enclosed Business Reply Form and fax back to 508-828-3762 or email to lregan3@its.jnj.com. For questions call Linda Regan @508-977-6606.