Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. 1) The Varian Low Energy Linear Accelerator is indicated for precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment i is indicated. (2) The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

This recall has been terminated (originally issued September 14, 2012).

Company
Varian Medical Systems, Inc. Oncology Systems
Recall Initiated
August 14, 2012
Posted
September 14, 2012
Terminated
December 4, 2013
Recall Number
Z-2383-2012
Quantity
221 Units
Firm Location
Palo Alto, CA
Official Source
View on FDA website ↗

Reason for Recall

Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume.

Distribution

Worldwide Distribution.

Lot / Code Info

LIST OF SERIAL NUMBERS:  H140863 ,H294608 ,H294669 ,H294732 ,H140865 ,H294610 ,H294670 ,H294733 ,H140866 ,H294611 ,H294671 ,H294734 ,H140867 ,H294612 ,H294672 ,H294735 ,H140868 ,H294613 ,H294673 ,H294736 ,H140869 ,H294614 ,H294674 ,H294737 ,H140870 ,H294615 ,H294675 ,H294738 ,H140871 ,H294616 ,H294676 ,H294739 ,H140872 ,H294617 ,H294677 ,H294740 ,H140873 ,H294618 ,H294678 ,H294741 ,H140874 ,H294619 ,H294679 ,H294742 ,H140875 ,H294620 ,H294680 ,H294743 ,H140876 ,H294621 ,H294681 ,H294744 ,H140877 ,H294622 ,H294682 ,H294745 ,H140878 ,H294623 ,H294683 ,H294746 ,H140879 ,H294624 ,H294684 ,H294747 ,H140880 ,H294625 ,H294685 ,H294748 ,H140881 ,H294626 ,H294686 ,H294749 ,H140882 ,H294627 ,H294687 ,H294750 ,H140883 ,H294628 ,H294688 ,H294751 ,H140884 ,H294629 ,H294689 ,H294752 ,H140885 ,H294630 ,H294690 ,H294753 ,H140886 ,H294631 ,H294691 ,H294754 ,H140887 ,H294632 ,H294692 ,H294755 ,H140888 ,H294633 ,H294693 ,H294756 ,H140889 ,H294634 ,H294694 ,H294757 ,H140890 ,H294635 ,H294695 ,H294758 ,H140891 ,H294636 ,H294696 ,H294759 ,H140892 ,H294637 ,H294697 ,H294760 ,H140893 ,H294638 ,H294698 ,H294761 ,H140895 ,H294639 ,H294699 ,H294762 ,H140896 ,H294640 ,H294700 ,H294763 ,H140897 ,H294641 ,H294701 ,H294764 ,H140898 ,H294642 ,H294702 ,H294765 ,H140899 ,H294643 ,H294704 ,H294766 ,H140900 ,H294644 ,H294705 ,H294767 ,H181175 ,H294645 ,H294706 ,H294768 ,H181176 ,H294646 ,H294707 ,H294769 ,H181177 ,H294647 ,H294708 ,H294770 ,H181178 ,H294648 ,H294709 ,H294771 ,H181179 ,H294649 ,H294710 ,H294772 ,H181180 ,H294650 ,H294711  ,H181181 ,H294651 ,H294712  ,H181182 ,H294652 ,H294713  ,H181183 ,H294653 ,H294714  ,H181184 ,H294654 ,H294715  ,H181185 ,H294655 ,H294716  ,H181186 ,H294656 ,H294717  ,H181187 ,H294657 ,H294718  ,H181188 ,H294658 ,H294719  ,H181189 ,H294659 ,H294720  ,H181190 ,H294660 ,H294722  ,H181191 ,H294661 ,H294723  ,H181192 ,H294662 ,H294724  ,H182000 ,H294663 ,H294726  ,H182001 ,H294664 ,H294727  ,H182002 ,H294665 ,H294728  ,H182003 ,H294666 ,H294729  ,H182005 ,H294667 ,H294730  ,H294607 ,H294668 ,H294731

Root Cause

Nonconforming Material/Component

Action Taken

Varian sent an "URGENT MEDICAL DEVICE CORRECTION URGENT FIELD SAFETY NOTICE" dated August 14, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-888-827-4265 for questions regarding this recall.