MLCi/MCLi2. Radiation therapy.

This recall has been terminated (originally issued May 23, 2012).

Company
Elekta, Inc.
Recall Initiated
April 19, 2012
Posted
May 23, 2012
Terminated
October 6, 2014
Recall Number
Z-1651-2012
Quantity
386
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

The Y2 diaphragm block was found loose.

Distribution

Nationwide Distribution including Puerto Rico.

Lot / Code Info

MRT 2424/MRT 14401

Root Cause

Nonconforming Material/Component

Action Taken

Elekta sent an "IMPORTANT FIELD SAFETY NOTICE" dated April 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to carry out the recommended weekly physics check of the X-ray and light field size calibration. If an error is found, customers are instructed to contact their local Elekta service engineer.