Non-DEHP Microbore Admin Set with Non-Vented Spike and Needleless Injection Port Positive Pressure Package Assembly, product code: 340-4133, Method of Sterilization using Ethylene Oxide. The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.
This recall has been terminated (originally issued May 30, 2012).
- Company
- Moog Medical Devices Group
- Recall Initiated
- May 18, 2012
- Posted
- May 30, 2012
- Terminated
- February 22, 2013
- Recall Number
- Z-1671-2012
- Quantity
- 3,880 sets.
- Firm Location
- Salt Lake City, UT
- Official Source
- View on FDA website ↗
Reason for Recall
Moog Medical Device Group became aware on March 23, 2012 via customer complaint, of a reversed pump segment on an Administration Set. This malfunction could reverse the flow of fluid or medication backwards from what was intended.
Distribution
Nationwide distribution: USA including states of: AZ, CA, FL, IL, KS, MA, MI, MD, NC, ND, NY, OK, OH, and TX.
Lot / Code Info
CF1129990
Root Cause
Process design
Action Taken
The firm, Moog Medical Device Group, sent an "Urgent Field Safety Notice" dated May 21, 2012 via certified letter and a press release direct to consignees/customers. The notice describes the product, problem and actions to be taken. The customers were instructed to do the following: If purchased by distributor -remove suspect product from inventory; contact your distributor to arrange replacement product and request a shipping return label as per the distributor's process and ship product to their location. If purchased by Moog-remove suspect product from inventory, contact Moog Customer Service at 1-800-970-2337, prompt #7 and Moog customer service will provide a call tag for convenient product return. For additional information or clarification, please contact the Post Market Surveillance Manager at (801) 264-1001, ext. 224 or via email at: lbrewer@moog.com.