S1827 Crea Calibration solution 1, 944-135, Lot DV-02 Radiometer Medical, Bronshoj Denmark Product Usage: The S1827 Crea Calibration Solution 1 (944-135) Lot DV-02 is a solution used by the ABL 8x7 series creatine analyzers to establish calibration points for measured parameters.

This recall has been terminated (originally issued February 13, 2012).

Company
Radiometer America Inc
Recall Initiated
December 16, 2011
Posted
February 13, 2012
Terminated
January 17, 2014
Recall Number
Z-1008-2012
Quantity
7 boxes/6/175mL bottles
Firm Location
Westlake, OH
Official Source
View on FDA website ↗

Reason for Recall

When the user activates the dosicap (containing the crea additive) and mounts it onto the bottle, the threads will not seal correctly causing approximately 10% of the solution to leak out when the bottle is placed horizontally for dissolving the additive. This will cause the concentration of creatinin and creatin in the calibration solution to be wrong, and results in a bias on the Crea results.

Distribution

Nationwide Distribution (USA) including the states of: NY, PA and TN.

Lot / Code Info

S1827 Part Number: 944-135, lot DV02

Root Cause

Device Design

Action Taken

Radiometer sent a Medical Device Recall letter dated December 28, 2011 to their customers. The letter provided details on the affected devices, description of the problem and actions to be taken. The firm instructed their customers to immediately examine their stock and return any remaining product to Raiometer America Inc. for disposal. The letter states that Radiometer will replace all bottles of the affected lots, at no charge. Customers were asked to complete the attached Recall Response Fax Form and fax to 800-736-0601. For questions regarding the Recall contact Radiometer America Technical support at 1-800-736-0600 opt. 4.