Merit Medical Systems, Inc., Convenience Trays, custom packaged under the following Merit Medical labels: (1) Prentice PICC Kit, Catalog number K12T-03698; (2) Insertion Kit, catalog number K12T-03050C; and (3) Angio Procedure Pack, catalog number K09T-09530A.

This recall has been terminated (originally issued November 23, 2009).

Company
Merit Medical Systems, Inc.
Recall Initiated
May 19, 2009
Posted
November 23, 2009
Terminated
December 3, 2009
Recall Number
Z-0413-2010
Quantity
378 units
Firm Location
Chester, VA
Official Source
View on FDA website ↗

Reason for Recall

Surgical gloves in hospital convenience kits may be degraded and tear during use.

Distribution

USA distribution: nine hospitals in Texas and Colorado.

Lot / Code Info

Lots: V705949, Exp 2010-05 V698581, Exp 2010-04 V700586, Exp 2012-04 V707921, Exp 2012-04

Root Cause

Nonconforming Material/Component

Action Taken

Merit Medical Systems notified customers by phone, fax, or visit by sales representatives on 05/19/09. The notification advised of the convenience kit correction due to the recall of the Derma Prene Ultra Surgical Gloves, and that use of the product should cease immediately. Sales representatives were instructed to provide replacement gloves and "warning" stickers to be placed on identified kits in inventory. The warning stickers advise the customer to discard the surgical gloves in the kit and replace. Customers were requested to verify product disposition by signing a response form.