AIMER ENDOFEMORAL 6MM OFFSET (GREEN) Part Number 72201716. The product is an orthopedic device.

This recall has been terminated (originally issued November 24, 2009).

Company
Smith & Nephew, Inc., Endoscopy Div.
Recall Initiated
September 3, 2009
Posted
November 24, 2009
Terminated
May 8, 2012
Recall Number
Z-0416-2010
Firm Location
Mansfield, MA
Official Source
View on FDA website ↗

Reason for Recall

Device may have improper Laser Marking, 4mm device may be marked as 6mm and 6mm device may be marked as 4mm.

Distribution

Worldwide Distribution -- United States including Puerto Rico, Australia, Austria, Belgium, Canada, Costa Rica, Dubai, Finland, France, Germany, Great Britain, Netherlands, Norway, Portugal, Spain, Sweden, and Switzerland.

Lot / Code Info

Lot Number 502780141

Root Cause

Labeling design

Action Taken

A notification letter was issued via Fedex on September 3, 2009 to domestic consignees and international consignees were notified via e-mail. The letter described the issue as the affected device may have improper laser marking. The 4mm devices are marked as 6mm and the 6mm devices are marked at 4mm. Customers were instructed to return the completed response form and affected product in inventory to Smith & Nephew. Direct questions about the recall to your local Smith & Nephew Sales Representatives or call 1-508-261-3655.