RecallDepth

Coloplast Titan OTR Inflatable Penile Prosthesis,18cm scrotal Bioflex cylinder set w/OTR Pump, P/N QSR9181400. Sterile EO. Manufacturer: Coloplast A/S, 3050 Denmark. Distributed by: Coloplast Corp., Minneapolis, MN 55411 USA. Indicated for male patients suffering from erectile dysfunction who are considered to be candidates for implantation of a penile prosthesis.

Company
Coloplast Manufacturing Us, Llc
Recall Initiated
October 6, 2009
Posted
November 10, 2009
Terminated
December 12, 2011
Recall Number
Z-0171-2010
Quantity
3
Firm Location
Minneapolis, MN

Reason for Recall

Coloplast determined that a portion of Titan OTR pumps were manufactured using a process that where validation activities were not completed. The Titan OTR pumps were made to our standard product specifications and have passed all in-process testing to ensure mechanical reliability of these pumps. Coloplast is not recommending product explant because there are no identified risks since the prod

Distribution

Nationwide Distribution -- FL, GA, NC.

Lot / Code Info

Lot 1919244.

Root Cause

Process control

Action Taken

Coloplast issued a customer notification letter dated October 2009 describing the affected product. For further information, contact Coloplast at 1-800-533-0464.