GE Healthcare MUSE Cardiology Information Systems vs. 7.0, 7.0.1, 7.0.2, 7.1, 7.1.1. The MUSE Cardiology Information System is intended to store, access and manage cardiovascular information on adult and pediatric patients.
This recall has been terminated (originally issued August 14, 2009).
- Company
- Ge Healthcare, Llc
- Recall Initiated
- February 2, 2009
- Posted
- August 14, 2009
- Terminated
- January 16, 2014
- Recall Number
- Z-1719-2009
- Quantity
- 327
- Firm Location
- Waukesha, WI
- Official Source
- View on FDA website ↗
Reason for Recall
GE Healthcare has become aware of an issue resulting from a workflow situation related to the automated labeling of patient data in the MUSE Cardiology Information System. When a prior record has been confirmed with an invalid (default) patient identifier (PID), MUSE will accept future records with the same default PID as correct. If subsequent records don not contain a name, MUSE will pre-popu
Distribution
Worldwide Distribution -- Including states of AL, AK, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, UT, VT, VA, WA, WI, WY and DC and countries of UNITED KINGDOM , TAIWAN, SWEDEN, SPAIN, SOUTH AFRICA, SINGAPORE, SAUDI ARABIA, QATAR, NEW ZEALAND, NETHERLANDS, KUWAIT, REPUBLIC OF KOREA, JORDAN, JAPAN, ITALY, IRELAND, GERMANY, FRANCE, FINLAND, ESTONIA, DENMARK, CANADA, BELGIUM, AUSTRIA and AUSTRALIA.
Lot / Code Info
Serial Numbers: SBP08365245GA, SBP08365257GA, SBP08124375GA, SBP07041811GA, SBP08164446GA, SBP08104129GA, SBP08104136GA, SBP08164458GA, SBP08385418GA, SBP07282702GA, SBP08124399GA, SBP07373018GA, SBP08194531GA, SBP08335087GA, SBP08335073GA, SBP07011728GA, SBP08164445GA, SBP07021744GA, SBP08365297GA, SBP05480036GA, SBP08335083GA, SBP08365284GA, SBP08365278GA, SBP08365282GA, SBP07423241GA, SBP08174469GA, SBP08194513GA, SBP08144422GA, SBP07463469GA, SBP07423262GA, SBP08365246GA, SBP06160272GA, SBP07463493GA, SBP07232542GA, SBP07182278GA, SBP07212445GA, SBP08124336GA, SBP08164453GA, SBP08144435GA, SBP08335106GA, SBP06300799GA, SBP08063821GA, SBP08023779GA, SBP08194514GA, SBP08335034GA, SBP07282642GA, SBP08294905GA, SBP08365260GA, SBP08114216GA, SBP07021746GA, SBP07232570GA, SBP08335088GA, SBP07081983GA, SBP08294892GA, SBP08104175GA, SBP08375355GA, SBP08184482GA, SBP08194534GA, SBP08244758GA, SBP08335108GA, SBP08114264GA, SBP08093988GA, SBP08335092GA, SBP07192318GA, SBP08375335GA, SBP07282681GA, SBP08335022GA, SBP08124374GA, SBP08104156GA, SBP07453378GA, SBP08194517GA, SBP08104154GA, SBP08204580GA, SBP08124395GA, SBP07282685GA, SBP08023778GA, SBP08335105GA, SBP08294899GA, SBP08355141GA, SBP07021745GA, SBP07082001GA, SBP08365244GA, SBP08365281GA, SBP08194523GA, SBP08224698GA, SBP08335090GA, SBP08355175GA, SBP08204613GA, SBP06330859GA, SBP08365270GA, SBP08053807GA, SBP08124365GA, SBP08365286GA, SBP08375351GA, SBP06441264GA, SBP08335107GA, SBP07061929GA, SBP08214644GA, SBP08114259GA, SBP08355167GA, SBP08194509GA, SBP08214646GA, SBP08124353GA, SBP07513715GA, SBP07021767GA, SBP08124385GA, SBP06330842GA, SBP08194506GA, SBP07252604GA, SBP07142196GA, SBP07312784GA, SBP08124383GA, SBP08385400GA, SBP08244734GA, SBP06451324GA, SBP08214658GA, SBP08335086GA, SBP08284827GA, SBP08124403GA, SBP06391110GA, SBP07182284GA, SBP08224699GA, SBP08244733GA, SBP08365202GA, SBP08124337GA, SBP08104105GA, SBP08104181GA, SBP07373052GA, SBP08184489GA, SBP08104149GA, SBP07503683GA, SBP08335109GA, SBP08335074GA, SBP08335091GA, SBP08104152GA, SBP08335030GA, SBP07413185GA, SBP08164463GA, SBP07373060GA, SBP08365276GA, SBP06280656GA, SBP08224701GA, SBP08114274GA, SBP08194535GA, SBP08405478GA, SBP08405471GA, SBP08385457GA, SBP08385462GA, SBP08385464GA, SBP08385446GA, SBP08385444GA, SBP08385443GA, SBP08385448GA, SBP08415533GA, SBP08415532GA, SBP08425546GA, SBP08425544GA, SBP08415499GA, SBP08425541GA, SBP08425555GA, SBP08425554GA, SBP08415496GA, SBP08415497GA, SBP08425553GA, SBP08415489GA, SBP08425550GA, SBP08415523GA, SBP08425569GA, SBP08415517GA, SBP08415515GA, SBP08415494GA, SBP08445598GA, SBP08425568GA, SBP08425570GA, SBP08425582GA, SBP08425567GA, SBP08445593GA, SBP08445611GA, SBP08455627GA, SBP08445608GA, SBP08455619GA, SBP08455643GA, SBP08475675GA, SBP08455634GA, SBP08475668GA, SBP08475663GA, SBP08475654GA, SBP08485711GA, SBP08475649GA, SBP08495730GA, SBP08475651GA, SBP08495752GA, SBP08495771GA, SBP08495750GA, SBP08505773GA, SBP08515836GA, SBP08505781GA, SBP08515840GA, SBP08515903GA, SBP08515900GA, SBP08495761GA, SBP08495756GA, SBP08515918GA, SBP08515925GA, SBP08515914GA, SBP08525975GA, SBP08525957GA, SBP08515948GA, SBP08525955GA, SBP09026018GA, SBP09026022GA, SBP08525963GA, SBP09026013GA, SBP08515947GA, SBP09036051GA, SBP08385461GA, SBP06521691GA, SBP05480037GA, SBP06140217GA / SBP08204594GA, SBP06160267GA, SBP06190362GA / SBP08204588GA, SBP07122132GA, SBP07403106GA, SBP07473532GA, SBP08094018GA, SBP06200394GA, 717933, 9011933, M2BV0021F, SBP06511552GA, SBP08294857GA, SBP08294933GA, SBR06120044GA, SBY07130180GA, SBY07360211GA, SBP06511555GA, SBP06340971GA, SBP06491460GA, C48R0044F, SBP07102027GA, SBP07122115GA, SBP07192300GA, SBP07122128GA, SBP06090174GA, SAC04380008GA, SAE05180093GA, SBP06371047GA, SBP07232538GA, SBP08194511GA, SBX06380017GA, SBX07220024GA, SBX08100045GA, SBY06230110GA, SBY06240113GA, SBY06470152GA, SBY06480154GA, SBY06490158GA, SBY07360210GA, SBP06230509GA, SBP06140216GA, SBP06190256GA, SBP06190332GA, SBP06210438GA, SBP07081986GA, SBP07122122GA, SBP07453464GA, SBP07483550GA, SBP07493604GA, SBP08104135GA, SBP08124342GA, SBP08365289GA, SBP08385409GA, SBP08475667GA, SBP08415529GA, SBP09026011GA, SBP06491456GA, SBP06491457GA, SBP06491464GA, SBP07282663GA, SBP07373044GA, SBP08375385GA, SBP08324972GA, SBP08335006GA, SBP08204620GA, SBP08415520GA, SBP06290743GA, SBP06501500GA, SBP08324976GA, SBP06180301GA, SBP06220456GA, SBP07041771GA, SBP07122125GA, SBP07413195GA, SBP08023771GA, SBP08104147GA, SBP08104150GA, SBP08485713GA, SBP06080168GA, SBP06371065GA, SBP06330868GA, SBP06451322GA, SBP06441295GA, SBP06160260GA, SBP07122130GA, SBP07061863GA, SBP07362961GA, SBP07423228GA, SBR07421143GA, SBP07403113GA, SBP08124390GA, SBP08093975GA, SBP08094067GA, SBP06090177GA, SAD05473249GA, SBP07092015GA, SBP08294935GA, SBY08260270GA, SBY07220206GA, SBP08415527GA, SBP06431234GA, SBP07332848GA, SBP08485718GA, SBX08150047GA and SBP08144418GA.
Root Cause
Software design
Action Taken
GE Healthcare issued an "Urgent Medical Device Correction" letter dated October 13, 2008, addressed to Healthcare Administrator/Risk Manager, Cardiology Administrators and Head of Cardiology. The letter listed the Safety Issue, Affected Product Details, Safety Instructions, Product Correction, Background Information and Contact Information. For further questions, contact GE Healthcare at 1-800-558-7044. Outside the US, please contact your local GE Healthcare Service Representative.