ACCELERATOR APS Centrifuge Module (Hettich Robotic centrifuge); List Number: 07L02-01; manufactured by Inpecco SpA, Segrate (MI), Italy.

This recall has been terminated (originally issued February 5, 2009).

Company
Abbott Laboratories, Inc
Recall Initiated
September 9, 2008
Posted
February 5, 2009
Terminated
October 4, 2010
Recall Number
Z-0951-2009
Quantity
6
Firm Location
Irving, TX
Official Source
View on FDA website ↗

Reason for Recall

ACCELERATOR APS Centrifuge Module being recalled due to potential for micro-cracks to form over time in the metal centrifuge rotor buckets that may cause the centrifuge buckets to disconnect from the rotor in the centrifuge during centrifugation.

Distribution

Nationwide Distribution --- including states of TX, UT, and VA.

Lot / Code Info

Product Control/Lot #SK28.99

Root Cause

Device Design

Action Taken

Notification letters sent to consignees on 09/09/08. Consignees provided instructions for inspecting the buckets for cracks and were asked to contact customer service reps for replacement buckets. Consignees also asked to respond via fax using Customer Reply form acknowledging receipt, understanding and completion of correction letter.