AMS InteXen LP Collagen Dermal Matrix, REF 72404066. Product Size: 8 x 12 cm. AMS InteXen LP is a sterile non-perforated processed porcine collagen dermal matrix. Sterile EO. American Medical Systems, Inc., 10700 Bren Road West, Minnetonka, MN 55343 USA. The AMS collagen dermal matric is intended for use in the treatment of hernias where connective tissue has ruptured or for implantation to reinforce soft tissues where weakness exists in the urological, gynecological, or gastroenterological anatomy.

Company
American Medical Systems
Recall Initiated
December 5, 2008
Posted
January 27, 2009
Terminated
June 5, 2012
Recall Number
Z-0932-2009
Quantity
5
Firm Location
W Minnetonka, MN

Reason for Recall

The product size shown on the carton labeling does not match the size on the product inside.

Distribution

Nationwide Distribution --- including states of TN and TX.

Lot / Code Info

lot/serial number: 567398002, 567398003

Root Cause

Labeling mix-ups

Action Taken

Customers were first notified by phone on 12/05/08. They were told of the situation and were requested to return the devices. Customers were also provided an Important Product Recall letter, dated 12/05/08. This letter identified affected product, told customers to discontinue use of product and to return it. An acknowledgement form was also asked to be returned. AMS Customer Service should be contacted at 1-800-328-3881 to receive a Return Authorization number and arrange for shipment of replacement product.