Equal Diagnostics Lipase Color Reagent Kit Part Number: 905-B For In Vitro Diagnostic Use Containing the Lipase Color Reagent, 30 ml, Part Number 80-2005-01.

Company
Genzyme Corporation
Recall Initiated
April 14, 2008
Posted
August 22, 2008
Terminated
September 16, 2008
Recall Number
Z-2148-2008
Quantity
3,443
Firm Location
Cambridge, MA

Reason for Recall

Calibration and Quality Control outside of established range resulting from an increase in initial OD readings with the reagent contained within the kits.

Distribution

Nationwide Distribution

Lot / Code Info

Reagent lot Number 491686, Lot #: 7EPM30 exp 01/15/09

Root Cause

Nonconforming Material/Component

Action Taken

Genzyme notified accounts by Fed'X on 4/14/08 via an Imprtant Field Correction letter, requesting users to discard kits and follow the "User Update" instructions previously provided to minimize calibration error. Customers are to complere and sign the Confirmation of Notification form and fax it back to the firm regardless if they have any affected product or not. Also, requires that the customer confirms that the REagent OD limits for First and Last H are set at 0.15 for Lipase on the Olympus AU400/400e, AU600, AU640/640e and AU2700/AU5400 Systems.