Zimmer Spine Cyclone anterior cervical plate, 4 Level, 64MM (Ti-6Al-4V); REF 600-04-164.

This recall has been terminated (originally issued September 1, 2008).

Company
Zimmer Inc.
Recall Initiated
March 26, 2008
Posted
September 1, 2008
Terminated
November 12, 2009
Recall Number
Z-1742-2008
Firm Location
Warsaw, IN
Official Source
View on FDA website ↗

Reason for Recall

The screw locking cap may fracture when the surgeon rotates it into the locked position.

Distribution

Arizona, California, Florida, Georgia, Louisiana, Michigan, Minnesota, Montana, New Hampshire, North Carolina, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, Washington and Wisconsin.

Lot / Code Info

All lots.

Root Cause

Device Design

Action Taken

Consignees were notified via voicemail on 3/26/08 and a letter was sent to all distributors on 3/27/08 informing them that the plate should no longer be used and that all lots are being recalled. A letter dated 6/18/08 was issued to implant physicians recommending patient monitoring. Contact Zimmer Spine at 1-952-857-5636 for assistance.