Triathlon Femoral Sizer - The Femoral sizer was developed for use with the Triathlon Total Knee Instrumentation system. It interfaces with the Modular Handled and the Femoral Stylus to facilitate sizing for a femoral component, and preparing holes to accept the corresponding 4:1 Cutting Guide. The femoral sizer is designed to prepare the femoral cuts in 3 degrees of external rotation by referencing against the posterior condyles. The Femoral sizer is set by locking it into the left or right orientation based on the femur.

This recall has been terminated (originally issued September 16, 2008).

Company
Stryker Howmedica Osteonics Corp.
Recall Initiated
May 27, 2005
Posted
September 16, 2008
Terminated
September 24, 2008
Recall Number
Z-1864-2008
Quantity
560 units
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

Stryker Orthopaedics became aware that when attempting to position the Triathlon Femoral Sizer, the pre-setting may slip from l (left knee) to R (right knee) and vice verse intraoperatively.

Distribution

To Stryker branches/agencies worldwide.

Lot / Code Info

Catalog Number: 6541-1-603 All lots (ecxept lot codes NMZC01, NMYV01, NMYW03.

Root Cause

Device Design

Action Taken

Letters were sent to all affected Stryker Branches/Agencies on May 27, 2005. Letters requested that each branch examine the inventory and hospitals locations to identify the instruments. Contact James Young at 201-831-5110, Brian Coughlin at 210-831-5665 or Rita Intorrella at 201-831-5825 if you have questions.