OEC Uroview 2800 with Dual Monitor Option, fluoroscopic x-ray system, Model Numbers: 882082-01, 882082-02, 886765-01, GE Healthcare, Surgery, Salt Lake City, UT.

This recall has been terminated (originally issued February 15, 2007).

Company
Ge Oec Medical Systems, Inc
Recall Initiated
January 15, 2007
Posted
February 15, 2007
Terminated
December 9, 2011
Recall Number
Z-0514-2007
Quantity
425 units
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Uroview 2800 system with dual monitors may pose a safety risk to patients if the monitors become disconnected and fall.

Distribution

Worldwide, including USA, Canada, China, France, Germany, Hong Kong, Italy, Japan, Singapore, South Africa, Switzerland, and Venezuela.

Lot / Code Info

Model Numbers: 882082-01, 882082-02, 886765-01. Units with manufactured dates between 03/2003 and 08/2006.

Root Cause

Other

Action Taken

Consignees were notified by letter on 01/15/2007. Instructions were given to position and not move the dual monitors until inspection and correction by GE personnel. For the affected units with a 3 phase power distribution box, users were cautioned to ensure that the access door to the power box is completely closed when the unit is powered on to prevent a shock hazard.