Coulter LH700 Series Hematology Analyzers, Part Number 6605632 with 2B Software (2B3, 2B4 and 2B5)

This recall has been terminated (originally issued March 6, 2007).

Company
Beckman Coulter Inc.
Recall Initiated
September 12, 2006
Posted
March 6, 2007
Terminated
April 19, 2012
Recall Number
Z-0566-2007
Quantity
1893 units in the US; 94 units to Canada
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

There is a potential for erroneous yet credible results for Body Fluids on the LH700 series hematology analyzers when a cassette label fails to read and the subsequent sample is cycled in Body Fluid mode.

Distribution

Nationwide and Canada

Lot / Code Info

Part Number 6605632, with 2B Software (2B3, 2B4 and 2B5)

Root Cause

Other

Action Taken

A Product Corrective Action (PCA) letter was mailed on Sep 12, 2006 to all LH 750 accounts to inform them that there is an issue with COULTER LH 700 Series analyzer software that could affect the Body Fluid analysis and that this this anomaly is present only in Software revisions of 2B. The letter instructs the customer on how to determine their software version. Immediate user preventive action was provided in the letter which is to reset the analyzer when a cassette label read error occurs. The letter advises that this information be shared with laboratory staff and be retained as part of their Quality System documentation. A customer response form was also included.