Capillary (end) caps, accessory for use with blood gas analyzers included in Radiometer''s CLINITUBES kits. The end caps are packaged in bags. These bags are placed in the capillary tube kit which also contains: capillary tubes, end caps, mixing wires, and a magnet. Each kit may contain 1 or 5 bags depending on the kit model. Each bag contains: 110
This recall has been terminated (originally issued August 3, 2006).
- Company
- Radiometer America Inc
- Recall Initiated
- March 22, 2006
- Posted
- August 3, 2006
- Terminated
- February 24, 2012
- Recall Number
- Z-1322-06
- Quantity
- Green Caps: 208 kits. Each kit contains 1 to 5 bags of caps.
- Firm Location
- Westlake, OH
- Official Source
- View on FDA website ↗
Reason for Recall
Capilliary end caps, an accessory for ABL blood gas analyzers, leak blood from capillary tubes during transport.
Distribution
Nationwide, Canada and Istanbul.
Lot / Code Info
Green Caps - Model #904-453 lot number less than 0535031
Root Cause
Other
Action Taken
The firm notified their customers of this recall action with a letter dated 3/22/2006. The letter was accompanied by a single bag of new end caps. The letter informs the customers of the ''leakage'' problem and asks that the customer examine their inventories and discard any of the referenced end caps found on-hand at any of their various locations. The letter also includes a response form to be faxed back to Radiometer indicating the amount of replacement capillary end caps required by the customer in addition to the one bag received. The recalling firm also notified the customers that the suspect end caps will all be replaced ASAP.