TC Monitor - Transcutaneous Monitor for p02 and pC02, Model #TCM4, (Cutaneous Oxygen Monitor)

This recall has been terminated (originally issued August 30, 2006).

Company
Radiometer America Inc
Recall Initiated
February 2, 2005
Posted
August 30, 2006
Terminated
May 9, 2012
Recall Number
Z-1437-06
Quantity
9 monitors were distributed.
Firm Location
Westlake, OH
Official Source
View on FDA website ↗

Reason for Recall

Alarm may be silenced- Rapid switching, on and off, of the TCM4 monitor may cause the built-in acoustical alarm to remain silent when tripped.

Distribution

Nationwide

Lot / Code Info

All Model #TCM4

Root Cause

Other

Action Taken

The firm initiated a recall action for the suspect device with a letter dated 2/2/2005. During a site visit to the U.S. customers conducted by the firm''s Field Service reps, the reps informed the user/customers of this problem and requested that all use of the device be discontinued, pending the formulation of a more permanent solution. Beginning on 2/8/2005, a second site visit was made to the same user/customers, at which time, the affected units were upgraded with operating software version 2.12. The firm''s recall/corrective action was scheduled for completion by 3/1/2005.